Left Heart Pacing—Experience with Several Types of Coronary Vein Leads
- 1 January 2002
- journal article
- Published by Springer Nature in Journal of Interventional Cardiac Electrophysiology
- Vol. 6 (1) , 71-75
- https://doi.org/10.1023/a:1014140716802
Abstract
Our experience with 121 coronary vein (CV) leads in 116 patients shows that CV leads are the leads of choice for pacing the left ventricle (LV). The information gained from pre-operative venous angiography permits individual selection of the most appropriate lead model for each case. The use of steerable electrophysiology catheters facilitates guide catheter cannulation of the coronary sinus (CS) when the anatomy is difficult and reduces the risk of complications. By selecting the CV lead model most suitable for each individual patient, we achieved successful implantation in 99.1% of patients. In this day and age, epicardial electrodes should be restricted to cases with CS anomalies which make CS cannulation impossible, and to LV lead implantation during heart surgery.Keywords
This publication has 6 references indexed in Scilit:
- Dissection of the coronary sinus secondary to pacemaker lead manipulation.Pacing and Clinical Electrophysiology, 2000
- Effect of Pacing Chamber and Atrioventricular Delay on Acute Systolic Function of Paced Patients With Congestive Heart FailureCirculation, 1999
- Improved Left Ventricular Mechanics From Acute VDD Pacing in Patients With Dilated Cardiomyopathy and Ventricular Conduction DelayCirculation, 1999
- Potential benefit of biventricular pacing in patients with congestive heart failure and ventricular tachyarrhythmiaThe American Journal of Cardiology, 1999
- Multisite Pacing as a Supplemental Treatment of Congestive Heart Failure: Preliminary Results of the Medtronic Inc. InSync StudyPacing and Clinical Electrophysiology, 1998