In Vitro Function and Durability Assessment of a Novel Polyurethane Heart Valve Prosthesis
- 1 September 1996
- journal article
- Published by Wiley in Artificial Organs
- Vol. 20 (9) , 1017-1025
- https://doi.org/10.1111/j.1525-1594.1996.tb04589.x
Abstract
While flexible‐leaflet, central‐flow prosthetic heart valves promise relief from anticoagulation therapy, they continue to be restricted by inadequate durability. In consequence, a novel trileaflet valve, made entirely from polyurethane, has been developed. A batch of 6 consecutively manufactured polyurethane valves was subjected to hydrodynamic function and accelerated fatigue testing. Computerized data acquisition and control systems have been introduced to improve valve testing methodologies. In terms of hydrodynamic function, the polyurethane valve demonstrates transvalvular pressure gradients similar to those for a bioprosthetic valve (Carpentier‐Edwards) and levels of retrograde flow significantly less than those for either the bioprosthetic valve or a bileaflet mechanical valve (St Jude Medical). The equivalent of 10 years of cycling without failure has been exceeded by all 6 polyurethane valves in accelerated fatigue tests with 2 valves remaining intact after 674 million cycles (equivalent to approximately 17 years) in continuing tests. Highspeed photography revealed considerable differences in leaflet motion between valves cycled at accelerated and physiological rates.Keywords
This publication has 23 references indexed in Scilit:
- Calcification and fatigue failure in a polyurethane heart valveBiomaterials, 1995
- A synthetic leaflet heart valve with improved opening characteristicsMedical Engineering & Physics, 1994
- Comparative study of the function of the Abiomed polyurethane heart valve for use in left ventricular assist devicesJournal of Biomedical Engineering, 1993
- Chapter 16 Prosthetic Cardiac ValvesCardiovascular Pathology, 1993
- Relative Blood Damage in the Three Phases of a Prosthetic Heart Valve Flow CycleAsaio Journal, 1993
- CAD-design, stress analysis and in vitro evaluation of three leaflet blood-pump valvesJournal of Biomedical Engineering, 1992
- Fabrication of a Jellyfish Valve for Use in an Artificial HeartAsaio Journal, 1992
- A synthetic three-leaflet valveJournal of Medical Engineering & Technology, 1992
- Hexsyn Trileaflet Valve: Application to Temporary Blood PumpsArtificial Organs, 1982
- The Design, Fabrication and Evaluation of a Trileaflet Prosthetic Heart ValveJournal of Biomechanical Engineering, 1980