Very High Pressure Gradient LC/MS/MS
- 10 May 2001
- journal article
- research article
- Published by American Chemical Society (ACS) in Analytical Chemistry
- Vol. 73 (13) , 2985-2991
- https://doi.org/10.1021/ac0010835
Abstract
A very high pressure liquid chromatography (VHPLC) system was constructed by modifying a commercially available pump in order to achieve pressures in excess of 1200 bar (17 500 psi). A computer-controlled low-pressure mixer was used to generate solvent gradients. Protein digests were rapidly analyzed by reversed-phase VHPLC with linear solvent gradients coupled to either a tandem mass spectrometer using electrospray ionization or a UV/visible detector. The separations were performed at pressures ranging from 790 (11 500 psi) to 930 bar (13 500 psi) in 22-cm-long capillary columns packed with C18-modified 1.5-μm nonporous silica particles. A digest of bovine serum albumin (BSA) was analyzed by the VHPLC system connected to a mass spectrometer in MS mode. An analysis of 12.5 fmol of sample gave signal-to-noise ratios of tryptic peaks greater than 10:1 in the base peak plot mass chromatogram. This system was also used to analyze a proteolytic digest of a rat liver protein excised from a 2-D gel separation of a liver tissue lysate. For this analysis, the mass spectrometer was set up to perform data-dependent scanning (automated switching from MS mode to MS/MS mode when a peak was detected) for peptide sequencing and protein identification by database searching. The results of this analysis are compared to an analysis performed on the same sample using the nanoelectrospray-MS/MS technique. Though both techniques were able to identify the unknown protein, the VHPLC method gave twice as many sequenced peptides as nanoelectrospray and improved the signal-to-noise ratio of the spectra by at least a factor of 10. Direct comparisons with nanoelectrospray for MS and MS/MS data acquisition from a BSA digest were made. These comparisons show enhancements of greater than 20-fold for VHPLC over nanoelectrospray. In addition, the VHPLC/MS/MS data acquisition was accomplished in an automated manner.Keywords
This publication has 8 references indexed in Scilit:
- MastheadAnalytical Chemistry, 1999
- Improvements in LC/Electrospray Ion Trap Mass Spectrometry Performance Using an Off-Axis NebulizerAnalytical Chemistry, 1999
- Ultrahigh-Pressure Reversed-Phase Liquid Chromatography in Packed Capillary ColumnsAnalytical Chemistry, 1997
- Femtomole sequencing of proteins from polyacrylamide gels by nano-electrospray mass spectrometryNature, 1996
- Enhanced sensitivity for peptide mapping with electrospray liquid chromatography-mass spectrometry in the presence of signal suppression due to trifluoroacetic acid-containing mobile phasesJournal of Chromatography A, 1995
- Ion spray mass spectrometric detection for liquid chromatography: A concentration- or a mass-flow-sensitive device?Journal of Chromatography A, 1993
- Nanoscale packed-capillary liquid chromatography coupled with mass spectrometry using a coaxial continuous-flow fast atom bombardment interfaceAnalytical Chemistry, 1991
- Preparation and evaluation of packed capillary liquid chromatography columns with inner diameters from 20 to 50 micrometersAnalytical Chemistry, 1989