Effect of Infrared Analyzer Homogenization Efficiency on Repeatability of Uncorrected Fat A and Fat B Signals

Abstract
Poor repeatability by infrared milk analyzers may be caused by inefficient homogenization as a result of light scattering and the Christiansen effect. The objectives of this study were to identify instruments with good and poor homogenization efficiency and to determine if a difference exists in repeatability performance between instruments with good vs poor homogenization efficiency. Unhomo-genized and homogenized portions of the same milk were tested 20 times consecutively on 22 instruments. An instrument was considered to have poor homogenization efficiency if the mean difference in the uncorrected signal between unhomo-genized and homogenized portions of the same milk was ≥1.43% of the fat test (i.e., ≥0.05% at 3.5% fat). Instruments were evaluated for repeatability by calculating the sample standard deviation and the range of the latter 19 uncorrected readings for un-homogenized and homogenized milks. When repeatability was evaluated as a function of homogenization efficiency, there was a significant (p = 0.001) correlation between poor homogenization efficiency and poor repeatability when testing unho-mogenized milk but not when testing homogenized milk. Improved homogenizer performance within infrared milk analyzers is needed to improve the repeatability of raw milk testing.