Biomechanical evaluation of Caspar and Cervical Spine Locking Plate systems in a cadaveric model

Abstract
✓ There exist two markedly different instrumentation systems for the anterior cervical spine: the Cervical Spine Locking Plate (CSLP) system, which uses unicortical screws with a locking hub mechanism for attachment, and the Caspar Trapezial Plate System, which is secured with unlocked bicortical screws. The biomechanical stability of these two systems was evaluated in a cadaveric model of complete C5–6 instability. The immediate stability was determined in six loading modalities: flexion, extension, right and left lateral bending, and right and left axial rotation. Biomechanical stability was reassessed following fatigue with 5000 cycles of flexion-extension, and finally, the spines were loaded in flexion until the instrumentation failed. The Caspar system stabilized significantly in flexion before (p < 0.05) but not after fatigue, and it stabilized significantly in extension before (p < 0.01) and after fatigue (p < 0.01). The CSLP system stabilized significantly in flexion before (p < 0.01) but not afte...