A quantitative assessment of portal film contrast as a function of beam energy
- 1 September 1986
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 13 (5) , 711-716
- https://doi.org/10.1118/1.595835
Abstract
Portal film contrast on a specially designed test phantom has been studied as a function of photon beam energy and object‐to‐film distance. The results provide important insights into the physical processes responsible for image contrast. In particular, theoretical calculations of Compton scatter reactions in the phantom can be used to predict visual film contrast. Good agreement between theory and experiment can be achieved by evaluating the double differential Compton cross sections [dσ(E,θ)/dE dθ] in the test object without resorting to variable parameters or artificial normalization. These calculations demonstrate the importance of low‐energy photons, object‐to‐film distance, and object size on portal film contrast.Funding Information
- U.S. Public Health Service (CA 38207)
- National Cancer Institute
This publication has 4 references indexed in Scilit:
- A digital image processing system for high energy x-ray portal imagesPhysics in Medicine & Biology, 1984
- Spectral characterization of 4 MV Bremsstrahlung by attenuation analysisMedical Physics, 1981
- Photon cross sections from 1 keV to 100 MeV for elements Z=1 to Z=100Atomic Data and Nuclear Data Tables, 1970
- X-Ray Spectra from Accelerators in the range 2 to 6 MeVPhysics in Medicine & Biology, 1967