An emittance–mass scanner for small-mass, low-energy beams
- 1 March 1996
- journal article
- research article
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 67 (3) , 1275-1276
- https://doi.org/10.1063/1.1146696
Abstract
An on‐line emittance–mass scanner (EMS), with a size of 10 cm×10 cm×8 cm, has been developed for use with low energy and light ion beams (A<40) at TRIUMF. Computerized data acquisition and procession give the rms emittances, intensities, and contour plots of each beam component. The angular resolution of the scanner is ±1.0 mrad, the maximum divergence of each component that the scanner can take is 140 mrad. Some experimental results of the measurement are given. The comparison between the contour shapes and the values of the emittance measured by both the emittance scanner and EMS is in good agreement.Keywords
This publication has 1 reference indexed in Scilit:
- An Emittance Scanner for Intense Low-Energy Ion BeamsIEEE Transactions on Nuclear Science, 1983