Response-time enhancement of a clinical gas analyzer facilitates measurement of breath-by-breath gas exchange
- 1 August 2000
- journal article
- research article
- Published by American Physiological Society in Journal of Applied Physiology
- Vol. 89 (2) , 581-589
- https://doi.org/10.1152/jappl.2000.89.2.581
Abstract
Tidal ventilation gas-exchange models in respiratory physiology and medicine not only require solution of mass balance equations breath-by-breath but also may require within-breath measurements, which are instantaneous functions of time. This demands a degree of temporal resolution and fidelity of integration of gas flow and concentration signals that cannot be provided by most clinical gas analyzers because of their slow response times. We have characterized the step responses of the Datex Ultima (Datex Instrumentation, Helsinki, Finland) gas analyzer to oxygen, carbon dioxide, and nitrous oxide in terms of a Gompertz four-parameter sigmoidal function. By inversion of this function, we were able to reduce the rise times for all these gases almost fivefold, and, by its application to real on-line respiratory gas signals, it is possible to achieve a performance comparable to the fastest mass spectrometers. With the use of this technique, measurements required for non-steady-state and tidal gas-exchange models can be made easily and reliably in the clinical setting.Keywords
This publication has 16 references indexed in Scilit:
- Pulmonary blood flow measured by inspiratory inert gas concentration forcing oscillationsRespiration Physiology, 1998
- A real-time algorithm to improve the response time of a clinical multigas analyser.Journal of Clinical Monitoring and Computing, 1998
- A457 ON-LINE GRAPHICAL TIME SYNCHRONIZATION PCO sub 2 AND AIRWAY FLOW TO DETERMINE CO sub 2 EXHALED PER BREATH (VCO sub 2, br)Anesthesiology, 1997
- Measurement of respiratory parameters by using inspired oxygen sinusoidal forcing signalsJournal of Applied Physiology, 1996
- Ventilatory inhomogeneity determined from multiple-breath washouts during sustained microgravity on Spacelab SLS-1Journal of Applied Physiology, 1995
- Correcting for the dynamic response of a respiratory mass spectrometerJournal of Applied Physiology, 1983
- Corrections for the response time and delay of mass spectrometersJournal of Applied Physiology, 1981
- Incorporating the gas analyzer response time in gas exchange computationsJournal of Applied Physiology, 1979
- Fast measurement of the CO2 partial pressure in gases and fluidsPflügers Archiv - European Journal of Physiology, 1978
- LUNG FUNCTION STUDIES. II. THE RESPIRATORY DEAD SPACEAmerican Journal of Physiology-Legacy Content, 1948