Accuracy of Area Measurements Made from MR Images Compared with Computed Tomography
- 1 January 1986
- journal article
- research article
- Published by Wolters Kluwer Health in Journal of Computer Assisted Tomography
- Vol. 10 (1) , 96-102
- https://doi.org/10.1097/00004728-198601000-00021
Abstract
The accuracy of area measurements made on magnetic resonance (MR) and CT images with clinical general purpose scanners was compared using automatic boundary detection. Both back projection and two-dimensional Fourier transform reconstruction were used with several pulse sequences for the MR images. Several phantoms were scanned with different receiver coils and image matrices. Errors caused by grey level variation in MR images can be reduced using local edge detection. The estimated values varied with the sequence, reconstruction algorithm, and the area under investigation. Magnetic resonance area measurements, should be interpreted with caution.This publication has 6 references indexed in Scilit:
- AN INVESTIGATION INTO THE PROBLEMS OF ATTENUATION AND AREA MEASUREMENTS MADE FROM CT IMAGES OF PULMONARY NODULES1984
- The assessment of achilles tendon size in primary hypercholesterolaemia by computed tomographyAtherosclerosis, 1982
- Volume determinations using computed tomographyAmerican Journal of Roentgenology, 1982
- Automatic Outlining of Regions on CT ScansJournal of Computer Assisted Tomography, 1981
- Determination of Liver, Kidney, and Spleen Volumes by Computed Tomography: An Experimental Study in DogsJournal of Computer Assisted Tomography, 1981
- Accurate Measurement of Liver, Kidney, and Spleen Volume and Mass by Computerized Axial TomographyAnnals of Internal Medicine, 1979