Assessment of coronary artery stenosis by magnetic resonance imaging.
Open Access
- 1 February 1996
- Vol. 75 (2) , 127-133
- https://doi.org/10.1136/hrt.75.2.127
Abstract
OBJECTIVE: The findings of magnetic resonance and x-ray angiography were compared for assessment of coronary artery stenosis in this validation study. BACKGROUND: Magnetic resonance angiography of the coronary arteries has recently been described, but there has been no comparison with x-ray angiography of localisation or assessment of important characteristics of coronary stenosis. METHODS: A breath hold, segmented k-space, 2D gradient echo imaging technique incorporating fat suppression was used in 39 patients (55 coronary stenoses) with known coronary artery disease. RESULTS: Overall, 47 stenoses (85%) were assessed by magnetic resonance (29 of 33 stenoses in the left anterior descending artery, one of one in the left main stem, 14 of 17 in the right coronary artery, and three of four in the left circumflex artery were detected). There was close agreement between magnetic resonance and x-ray angiography for the distance of the stenosis from the arterial origin (magnetic resonance mean (SD) 27 (16) mm versus x-ray angiography 27 (16) mm, P = NS, mean difference -0.2 mm). The distance to 39 stenoses (83%) agreed to within 5 mm, with increased scatter for more distal stenoses. The severity of magnetic resonance signal loss, assessed visually at the site of stenosis, varied significantly according to the percentage diameter stenosis (F = 30, P < 0.0001); stenosis severity with severe signal loss was 89 (7)%, with partial signal was 70 (16)%, and with irregular wall only 37 (11)%, with significant differences among the three groups (P < 0.001). A significant correlation was found between the proportional magnetic resonance signal loss at the stenosis and the percentage diameter stenosis severity (r = -0.67, P < 0.0001). The length of stenosis measured by magnetic resonance (6 (3) mm) was greater than by x-ray angiography (5 (2) mm, P < 0.006, mean difference +1.1 mm). Spearman's rank test showed that there was significant overestimation of stenosis length by magnetic resonance as stenosis severity increased (rs = 0.34, P < 0.02). CONCLUSIONS: Accurate localisation of coronary stenosis and a qualitative assessment of stenosis severity are possible by magnetic resonance, but stenosis length is overestimated as severity increases, probably because of disturbed patterns of flow with turbulence distal to severe stenoses. Reasonable results for the detection of coronary artery stenosis by magnetic resonance were achieved in this highly selected population, but further progress in imaging techniques is necessary before moving towards appreciable clinical application.Keywords
This publication has 17 references indexed in Scilit:
- Dendrimer‐based metal chelates: A new class of magnetic resonance imaging contrast agentsMagnetic Resonance in Medicine, 1994
- Magnetic resonance imaging of coronary arteries: technique and preliminary results.Heart, 1993
- A monitoring, feedback, and triggering system for reproducible breath‐hold MR imagingMagnetic Resonance in Medicine, 1993
- A Preliminary Report Comparing Magnetic Resonance Coronary Angiography with Conventional AngiographyNew England Journal of Medicine, 1993
- Fast Spiral Coronary Artery ImagingMagnetic Resonance in Medicine, 1992
- Interobserver variability in grading of coronary arterial narrowings using the American College of Cardiology/American Heart Association grading criteriaThe American Journal of Cardiology, 1992
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986
- Does Visual Interpretation of the Coronary Arteriogram Predict the Physiologic Importance of a Coronary Stenosis?New England Journal of Medicine, 1984
- Mortality related to cardiac catheterization and angiographyCatheterization and Cardiovascular Diagnosis, 1982
- Complications of coronary arteriography from the Collaborative Study of Coronary Artery Surgery (CASS).Circulation, 1979