Rapid Method for the Measurement of Monochromator Parameters in the near Infrared
- 1 March 1979
- journal article
- research article
- Published by SAGE Publications in Applied Spectroscopy
- Vol. 33 (2) , 105-107
- https://doi.org/10.1366/0003702794925994
Abstract
This paper describes a technique used for measuring monochromator characteristics in the near IR. It is shown that with an interferometer spectrophotometer, one can obtain parameters such as stray light, efficiency, resolution, bandwidth, and accuracy of wavelength readout. The advantage of this technique is simplicity and speed in data gathering.Keywords
This publication has 3 references indexed in Scilit:
- High-energy spectrophotometer for rapid constituent analysis in the range of 0.25–2.4 μmReview of Scientific Instruments, 1979
- Fourier Transform Infrared SpectroscopyScience, 1976
- Interferometry in the SeventiesAnalytical Chemistry, 1974