Blood pressure measurement during transport A comparison of direct and oscillotonometric readings in critically ill patients
- 1 August 1990
- journal article
- research article
- Published by Wiley in Anaesthesia
- Vol. 45 (8) , 659-665
- https://doi.org/10.1111/j.1365-2044.1990.tb14393.x
Abstract
This study assessed the accuracy of oscillotonometric blood pressure measurement in critically ill patients during transfer. Direct intra-arterial pressure measurements were taken in 44 transported patients as a 'gold standard' and compared with readings from four portable automatic oscillotonometers--the Dinamap 8100, Lifestat 100, Propaq 102 and Takeda UA711. All under-read systolic pressure (by 13%, 21%, 19% and 13% respectively) and over-read diastolic pressure (by 15%, 5%, 27% and 15% respectively) in comparison to direct pressure measurement. The limits of agreement for 95% of comparisons were broad. The systematic difference between direct and oscillotonometric measurements and the variability between and within patients show that direct oscillotonometric measurements are not interchangeable in these patients. Portable automatic oscillotonometers should not be substituted for direct monitoring in such circumstances.Keywords
This publication has 23 references indexed in Scilit:
- Accuracy of blood pressure measurements made aboard helicoptersAnnals of Emergency Medicine, 1988
- Transporting critically ill patients by ambulance: audit by sickness scoringBMJ, 1988
- Comparison of Direct Blood Pressure Measurements at the Radial and Dorsalis Pedis Arteries during Sodium Nitroprusside- and Isoflurane-induced HypotensionAnesthesiology, 1986
- Oscillometry and direct blood pressure: A comparative clinical study during deliberate hypotensionCanadian Journal of Anesthesia/Journal canadien d'anesthésie, 1986
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986
- Automatic blood pressure monitors A clinical evaluation of five models in adultsAnaesthesia, 1985
- An assessment of the Dinamap 845Anaesthesia, 1984
- Direct Blood Pressure Measurement —Dynamic Response RequirementsAnesthesiology, 1981
- Systolic Pressure Amplification in the Arteries of Normal SubjectsScandinavian Journal of Clinical and Laboratory Investigation, 1974
- Disparities Between Aortic and Peripheral Pulse Pressures Induced by Upright Exercise and Vasomotor Changes in ManCirculation, 1968