Collagen fiber disruption occurs independent of calcification in clinically explanted bioprosthetic heart valves
- 6 August 2002
- journal article
- research article
- Published by Wiley in Journal of Biomedical Materials Research
- Vol. 62 (3) , 359-371
- https://doi.org/10.1002/jbm.10293
Abstract
The durability of bioprosthetic heart valves (BHV) is severely limited by tissue deterioration, manifested as calcification and mechanical damage to the extracellular matrix. Extensive research on mineralization mechanisms has led to prevention strategies, but little work has been done on understanding the mechanisms of noncalcific matrix damage. The present study tested the hypothesis that calcification-independent damage to the valvular structural matrix mediated by mechanical factors occurs in clinical implants and could contribute to porcine aortic BHV structural failure. We correlated quantitative assessment of collagen fiber orientation and structural integrity by small angle light scattering (SALS) with morphologic analysis in 14 porcine aortic valve bioprostheses removed from patients for structural deterioration following 5–20 years of function. Calcification of the explants varied from 0 (none) to 1+ (minimal) to 4+ (extensive), as assessed radiographically. SALS tests were performed over entire excised cusps using a 0.254-mm spaced grid, and the resultant structural information used to generate maps of the local collagen fiber damage that were compared with sites of calcific deposits. All 42 cusps showed clear evidence of substantial noncalcific structural damage. In 29 cusps that were calcified, structural damage was consistently spatially distinct from the calcification deposits, generally in a distribution similar to that noted in porcine BHV subjected to in vitro durability testing. Our results suggest a mechanism of noncalcific degradation dependent on cuspal mechanics that could contribute to porcine aortic BHV failure. © 2002 Wiley Periodicals, Inc. J Biomed Mater Res 62: 359–371, 2002Keywords
This publication has 33 references indexed in Scilit:
- Primary Tissue Failure of Bioprostheses: New Evidence from In Vitro Tests*The Thoracic and Cardiovascular Surgeon, 2001
- Mechanisms of bioprosthetic heart valve failure: Fatigue causes collagen denaturation and glycosaminoglycan lossJournal of Biomedical Materials Research, 1999
- Effects of Mechanical Fatigue on the Bending Properties of the Porcine Bioprosthetic Heart ValveAsaio Journal, 1999
- The aortic valve microstructure: Effects of transvalvular pressureJournal of Biomedical Materials Research, 1998
- Pathology of Substitute Heart Valves: New Concepts and DevelopmentsJournal of Cardiac Surgery, 1994
- Biomechanical and structural properties of the explanted bioprosthetic valve leafletsJournal of Biomechanics, 1994
- Detection of remnant proteolytic activities in unimplanted glutaraldehyde‐treated bovine pericardium and explanted cardiac bioprosthesesJournal of Biomedical Materials Research, 1993
- Tissue Buckling as a Mechanism of Bioprosthetic Valve FailureThe Annals of Thoracic Surgery, 1988
- An ‘in vitro’ study of mechanical fatigue in glutaraldehyde-treated porcine aortic valve tissueBiomaterials, 1980
- Structural changes in glutaraldehyde-treated porcine heterografts used as substitute cardiac valves: Transmission and scanning electron microscopic observations in 12 patientsThe American Journal of Cardiology, 1978