A Third Generation Water Bath Based Blackbody Source
Open Access
- 1 September 1995
- journal article
- Published by National Institute of Standards and Technology (NIST) in Journal of Research of the National Institute of Standards and Technology
- Vol. 100 (5) , 591-599
- https://doi.org/10.6028/jres.100.044
Abstract
A third generation water bath based black-body source has been designed and constructed in the Radiometric Physics Division at the National Institute of Standards and Technology, Gaithersburg, MD. The goal of this work was to design a large aperture blackbody source with improved temporal stability and reproducibility compared with earlier designs, as well as improved ease of use. These blackbody sources operate in the 278 K to 353 K range with water temperature combined standard uncertainties of 3.5 mK to 7.8 mK. The calculated emissivity of these sources is 0.9997 with a relative standard uncertainty of 0.0003. With a 50 mm limiting aperture at the cavity; entrance, the emissivity increases to 0.99997.Keywords
This publication has 3 references indexed in Scilit:
- Theoretical Analysis of Laboratory Blackbodies 1: A Generalized Integral EquationApplied Optics, 1973
- Note on the Quality of Freezing Point BlackbodiesApplied Optics, 1971