Measurement of Inverted Electrical Conductivity Profiles Using Electromagnetic Induction

Abstract
Due to sharp discrepancies in the relative response of the horizontal and vertical coil configurations of the Geonics Limited EM‐38 electromagnetic induction soil conductivity meter for the top 15 cm of soil, the accuracy of measurement of bulk electrical conductivity for soil having a high surface electrical conductivity relative to deeper depths (i.e., an inverted electrical conductivity profile) has been unreliable. Two new approaches have been developed to compensate for these discrepancies, either through compensations in the vertical coil configuration response curve or the reestablishment of EM0,H adjustment curves utilizing data solely from inverted conductivity profiles. Both approaches yield more consistently reliable calculated bulk soil electrical conductivities when compared to measured electrical conductivities using the four‐electrode probe. The latter approach, however, appears to be more accurate.