Contact pressures in the human hip joint measured in vivo.
- 1 May 1986
- journal article
- research article
- Published by Proceedings of the National Academy of Sciences in Proceedings of the National Academy of Sciences
- Vol. 83 (9) , 2879-2883
- https://doi.org/10.1073/pnas.83.9.2879
Abstract
The pressures on human articular cartilage have been measured in vivo. An instrumented femoral head prosthesis that telemeters interarticular pressure at 10 discrete locations 253 times per second was implanted in apposition to natural acetabular cartilage. Data were acquired during surgery, recovery, rehabilitation, and normal activity, for longer than 1 year after surgery. Pressure magnitudes were synchronized with body-segment kinematic data and foot-floor force measurements so as to locate transduced pressure areas on the natural acetabulum and correlate movement kinematics and dynamics with local cartilage pressures. The data reveal very high local (up to 18 MPa) and nonuniform pressures, with abrupt spatial and temporal gradients, that correlate well both in magnitude and distribution with in vitro data and computer simulations of synovial joint mechanics. Peak pressures in vivo are, however, considerably higher than pressures measured in vitro under the putative forces experienced by the joint in life, particularly in normal movements where cocontraction occurs in agonist and antagonist muscles across the hip joint. Thus, extant gait-analysis studies which apply inverse Newtonian calculations to infer joint forces establish the lower limit on such forces, since such analyses include only the net muscular torques about the joint and cannot account for the contribution of the increment in joint force due to muscular cocontraction. Our data also contribute to the understanding of normal synovial joint tribology and the possible role of mechanical factors in the deterioration evident in osteoarthritis. Further, design criteria for both partial and total hip replacement prostheses and specific aspects of rehabilitation protocols following hip surgery (e.g., the extent to which crutches and canes unload the hip joint) warrant reconsideration in light of the extraordinary high pressures measured during the activities of daily living.This publication has 19 references indexed in Scilit:
- The frequency content of gaitPublished by Elsevier ,2004
- Singular Perturbation Analysis of the Nonlinear, Flow-Dependent Compressive Stress Relaxation Behavior of Articular CartilageJournal of Biomechanical Engineering, 1985
- A technique for measuring instantaneous in vitro contact stress distributions in articular jointsJournal of Biomechanics, 1982
- Degenerative Lumbar Spinal Stenosis:Results of Operative TreatmentActa Orthopaedica, 1982
- An evaluation of the approaches of optimization models in the prediction of muscle forces during human gaitJournal of Biomechanics, 1981
- Improved techniques for measuringin vitro the geometry and pressure distribution in the human acetabulum—II. Instrumented endoprosthesis measurement of articular surface pressure distributionJournal of Biomechanics, 1981
- A model for surface flow in cartilageJournal of Biomechanics, 1980
- Influence of Cartilage Geometry on the Pressure Distribution in the Human Hip JointScience, 1979
- In vivo records of hip loads using a femoral implant with telemetric output (a prelimary report)Journal of Biomedical Engineering, 1979
- The transmission of load through the human hip jointJournal of Biomechanics, 1971