Effect of Dialysate Temperature and Flow Rate on Peritoneal Clearance
- 23 October 1967
- journal article
- research article
- Published by American Medical Association (AMA) in JAMA
- Vol. 202 (4) , 363-365
- https://doi.org/10.1001/jama.1967.03130170163035
Abstract
Intermittent peritoneal dialysis was performed to observe the effect of dialysate temperature and flow rate on peritoneal clearance. At the same hourly flow rates, the dialysate heated to 37 C had a 35% greater clearance than the dialysate at 20 C. The most efficient method of dialysis with commercially available equipment is a repeated 2-liter exchange of dialysate at 37 C with no pause between inflow and outflow. This method in an average dialysate volume of 3.3 liters/hr and an average urea nitrogen clearance of 20.8 ml/min.Keywords