Heterogeneous enzyme immunoassay with electrochemical detection: competitive and "sandwich"-type immunoassays.
Open Access
- 1 September 1985
- journal article
- research article
- Published by Oxford University Press (OUP) in Clinical Chemistry
- Vol. 31 (9) , 1546-1549
- https://doi.org/10.1093/clinchem/31.9.1546
Abstract
In these competitive and "sandwich"-type heterogeneous enzyme immunoassays, based on liquid chromatography with electrochemical detection, rabbit immunoglobulin G is used as a model compound. Alkaline phosphatase (EC 3.1.3.1), the labeling enzyme, catalyzes conversion of phenyl phosphate to phenol. After separation on an octyldecylsilane column, the enzyme-generated phenol is detected in a thin-layer cell at a carbon-paste working electrode. The detection limit for phenol is 5.0 nmol/L. The electrode response varies linearly with concentration over a range of three orders of magnitude. For the sandwich-type assay procedure the detection limit is 10 ng/L; the linearity ranges over four orders of magnitude. The detection limit of the competitive immunoassay is 5 micrograms/L. The dynamic range spans two orders of magnitude.This publication has 4 references indexed in Scilit:
- Enzyme immunoassay for gentamicin.Clinical Chemistry, 1978
- Enzyme immunoassays with special reference to ELISA techniques.Journal of Clinical Pathology, 1978
- Quantitative enzyme immunoassay: current status.Clinical Chemistry, 1976
- ENZYME-IMMUNOASSAY1976