Tachycardia Mediated by an AV Universal (DDD) Pacemaker Triggered by a Ventricular Depolarization
- 1 January 1986
- journal article
- case report
- Published by Wiley in Pacing and Clinical Electrophysiology
- Vol. 9 (1) , 105-107
- https://doi.org/10.1111/j.1540-8159.1986.tb05366.x
Abstract
A 44‐year‐old female had Chagas’ disease with right bundle branch block, right axis deviation, and ventricular ectopic beats. An invasive evaluation showed a prolonged HV interval of 100 ms. A Medtronic Versatrax 7000 was implanted, after which a pacemaker‐mediated tachycardia was observed. The esophageal lead ruled out the possibility of an atrial triggered tachycardia. The diagnosis was pacemaker‐mediated tachycardia due to abnormal sensing of the terminal forces of the ventricular activation because of malposition of the atrial lead in the right ventricular out‐flow tract.Keywords
This publication has 3 references indexed in Scilit:
- Endless Loop Tachycardia in an AV Universal (DDD) PacemakerPacing and Clinical Electrophysiology, 1982
- Pacemaker Related TachycardiasPacing and Clinical Electrophysiology, 1982