Dose rate distribution around60Co,137Cs,198Au,192Ir,241Am,125I (models 6702 and 6711) brachytherapy sources and the nuclide99Tcm
- 1 October 1992
- journal article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 37 (10) , 1859-1872
- https://doi.org/10.1088/0031-9155/37/10/004
Abstract
Simple analytical functions derived from our point source Monte Carlo calculations on the combined attenuation and scatter factor, B exp(-mu r), for 60Co, 137Cs, 198Au, 192Ir, 241Am, 125I (models 6702 and 6711) brachytherapy sources and the nuclide 99Tcm, for water spherical geometries of radii R = 15 and 20 cm, are presented. Our results for the broadly used 60Co, 137Cs, 198Au and 192Ir brachytherapy sources can be compared directly and found in excellent agreement with the widely accepted data of Meisberger et al in the limited distance range for which the latter are valid. Our data, however, can be used with high accuracy outside this distance range. Many discrepancies observed among different data sets available in recent literature are attributable to differences in geometries used. The results for the recently introduced 241Am source are very dissimilar to those produced by any other currently used brachytherapy source. Dose rate distributions, based on the above simple functions, are proposed in accordance with the recommendations for calibration of the brachytherapy sources in terms of reference air kerma rate and were found to be in good agreement with data available in the literature. Our calculations for 125I sources (models 6702 and 6711), provided that the characteristic x-rays from titanium encapsulation are taken into account, support recent experimental and theoretical dose rate distributions indicating that currently accepted values for 125I may be overestimated.Keywords
This publication has 30 references indexed in Scilit:
- Photon mass attenuation and energy-absorption coefficientsPublished by Elsevier ,2002
- Dose rate determination for 125I seedsMedical Physics, 1990
- Dose distribution of 1 2 5 I sources in different tissuesMedical Physics, 1990
- The accuracy of single-seed dose superposition for I-125 implantsMedical Physics, 1989
- Monte Carlo calculation of energy build-up factors in the range from 15 keV to 100 keV, with special reference to the dosimetry of125I seedsPhysics in Medicine & Biology, 1988
- Revisions to radial dose function data for 1 2 5 I and 1 3 1 CsMedical Physics, 1986
- Physical characteristics of scattered radiation in diagnostic radiology: Monte Carlo simulation studiesMedical Physics, 1985
- Specification of brachytherapy sourcesThe British Journal of Radiology, 1984
- Some theoretical derivations relating to the tissue dosimetry of brachytherapy nuclides, with particular reference to iodine-125Medical Physics, 1983
- A Monte Carlo derivation of parameters for use in the tissue dosimetry of medium and low energy nuclidesThe British Journal of Radiology, 1982