Chronic intracortical microstimulation (ICMS) of cat sensory cortex using the Utah intracortical electrode array
- 1 March 1999
- journal article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Rehabilitation Engineering
- Vol. 7 (1) , 56-68
- https://doi.org/10.1109/86.750552
Abstract
In an effort to assess the safety and efficacy of focal intracortical microstimulation (ICMS) of cerebral cortex with an array of penetrating electrodes as might be applied to a neuroprosthetic device to aid the deaf or blind, we have chronically implanted three trained cats in primary auditory cortex with the 100-electrode Utah Intracortical Electrode Array (UIEA). Eleven of the 100 electrodes were hard-wired to a percutaneous connector for chronic access. Prior to implant, cats were trained to "lever-press" in response to pure tone auditory stimulation. After implant, this behavior was transferred to "lever-presses" in response to current injections via single electrodes of the implanted arrays. Psychometric function curves relating injected charge level to the probability of response were obtained for stimulation of 22 separate electrodes in the three implanted cats. The average threshold charge/phase required for electrical stimulus detection in each cat was, 8.5, 8.6, and 11.6 nC/phase respectively, with a maximum charge/phase of 26 nC/phase and a minimum of 1.5 nC/phase thresholds were tracked for varying time intervals, and seven electrodes from two cats were tracked for up to 100 days. Electrodes were stimulated for no more than a few minutes each day. Neural recordings taken from the same electrodes before and after multiple electrical stimulation sessions were very similar in signal/noise ratio and in the number of recordable units, suggesting that the range of electrical stimulation levels used did not damage neurons in the vicinity of the electrodes. Although a few early implants failed, we conclude that ICMS of cerebral cortex to evoke a behavioral response can be achieved with the penetrating UIEA. Further experiments in support of a sensory cortical prosthesis based on ICMS are warranted.Keywords
This publication has 27 references indexed in Scilit:
- The Utah Intracortical Electrode Array: A recording structure for potential brain-computer interfacesPublished by Elsevier ,1999
- Feasibility of a visual prosthesis for the blind based on intracortical micro stimulation of the visual cortexBrain, 1996
- Biocompatibility of silicon‐based electrode arrays implanted in feline cortical tissueJournal of Biomedical Materials Research, 1993
- Repetitive Microstimulation Alters the Cortical Representation of Movements in Adult RatsSomatosensory & Motor Research, 1990
- Neuronal activity evoked by chronically implanted intracortical microelectrodesExperimental Neurology, 1986
- Excitation of visual cortex neurons by local intracortical microstimulationExperimental Neurology, 1983
- Structure of human auditory cortex. I. Cytoarchitectonics and dendritic distributionsBrain Research, 1981
- Histological Evaluation of Neural Damage from Electrical StimulationNeurosurgery, 1981
- Experimental Nondestructive Electrical Stimulation of the Brain and Spinal CordJournal of Neurosurgery, 1970
- Excitation and accommodation in nerveProceedings of the Royal Society of London. B. Biological Sciences, 1936