Design Studtes of Two Possible Detector Blocks for High Resolution Positron Emission Tomography of the Brain
- 1 January 1987
- journal article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Nuclear Science
- Vol. 34 (1) , 344-348
- https://doi.org/10.1109/TNS.1987.4337360
Abstract
Two possible detector designs for high resolution positron camera systems have been investigated. The goal is to achieve an instrument that can measure the whole brain with a spatial resolution of 5 mm FWHM in all directions. For both detectors BGO crystals are used, with the dimension 4.5*9.5*25 mm. One detector scheme utilizes the Anger principle for crystal identification with 16 crystals mounted on two dual PMT:s via a 3 mm light guide. The other detector scheme utilizes position sensitive PMT:s. The figures of merit for these two configurations are discussed in terms of high count rate capabilities and identification reliability.Keywords
This publication has 2 references indexed in Scilit:
- Figures of Merit for Different Detector (Configurations Utilized in High Resolution Positron CamerasIEEE Transactions on Nuclear Science, 1986
- Analytical Study of the Performance of a Multilayer Positron Computed Tomography ScannerJournal of Computer Assisted Tomography, 1982