Noninvasive ventilatory monitoring with bellows pneumographs in supine subjects
- 1 August 1983
- journal article
- research article
- Published by American Physiological Society in Journal of Applied Physiology
- Vol. 55 (2) , 598-606
- https://doi.org/10.1152/jappl.1983.55.2.598
Abstract
We compared simultaneous measurements of ventilatory movements obtained by a bellows pneumograph (BP) and a respiratory inductive plethysmograph (RIP) vs. integrated volume from a pneumotachograph in seven healthy volunteers during different respiratory patterns. The purpose of this study was to assess if a computer-aided calibration procedure could improve the accuracy of BP, a simple semiquantitative method for noninvasive ventilatory monitoring in supine subjects. Both devices were repeatedly calibrated against a pneumotachograph in the same posture with a computer-aided least-squares method. One calibration maneuver was sufficient to achieve a minimal relative difference in volume measurements between the RIP as well as the BP and the pneumotachograph of less than 1.2 +/- 4.5 (SD) %. The accuracy remained in this range during the subsequent calibrations with time (1 h) and after body movements. However, this difference increased significantly with both devices when the subjects were studied in the lateral decubitus position. The present study indicates that despite theoretical advantages of the RIP, the BP, when properly calibrated, has similar performances in supine subjects for monitoring ventilation; it is simpler and less expensive than the RIP and is devoid of electrical drift and artifacts.This publication has 8 references indexed in Scilit:
- VALIDATION OF RESPIRATORY INDUCTIVE PLETHYSMOGRAPHY USING DIFFERENT CALIBRATION PROCEDURESPublished by Elsevier ,1982
- Noninvasive ventilatory monitoring by respiratory inductive plethysmography in conscious sheepJournal of Applied Physiology, 1981
- Changes in ventilation and chest wall mechanics during sleep in normal adolescentsJournal of Applied Physiology, 1981
- RESPIRATORY INDUCTION PLETHYSMOGRAPHY (RESPITRACE) - AN EVALUATION OF ITS USE IN THE INFANTPublished by Elsevier ,1981
- CHEST WALL MECHANICS AND PATTERN OF BREATHING DURING SLEEP IN ASTHMATIC ADOLESCENTSPublished by Elsevier ,1981
- Effects of respiratory apparatus on breathing patternJournal of Applied Physiology, 1980
- NON-INVASIVE MEASUREMENT OF VENTILATION DURING EXERCISE USING A RESPIRATORY INDUCTIVE PLETHYSMOGRAPH .1.Published by Elsevier ,1980
- Measurement of the separate volume changes of rib cage and abdomen during breathingJournal of Applied Physiology, 1967