Abstract
The pH is one of the most important parameters for char-acterizing the chemical properties of an aqueous solution. The glass eiectrode is by far the most commonly used pH sensor. The determination of pH in special situations, e.g., in vivo applications where the fragility of the glass electrode is a draw-back, requires pH sensors that can easily be miniaturized and built into physically rugged sleeves. Also, for other applications where the volume of solution is very restricted miniaturization of pH sensors is very important. The glass electrode does not respond properly to pH in some corrosive environments (e.g., in hydrogen fluoride [HF] solutions). At present, among alternatives to the hydrogen-selective glass electrode group of sensors, potentiometric metal-metal oxide pH sensors respond to pH, ideally due to an equilibrium involving the metal and its oxide. In the case of metal oxide electrodes, the metal is not involved in the potential-determining reaction. This distinction is of course not clear in many cases because the mechanism giving the pH response is not always unequivocal.Keywords
This publication has 115 references indexed in Scilit:
- Proteolytic enzyme modified metal oxide electrodes as potentiometric sensorsAnalytical Chemistry, 1986
- Physicochemical characterization of Co3O4 prepared by thermal decomposition II: Response to solution pHSurface Technology, 1984
- A multipoint micro antimony pH electrode for tissue surface measurementsJournal of Clinical Monitoring and Computing, 1984
- Surface and bulk processes at oxidized iridium electrodes—II. Conductivity-switched behaviour of thick oxide filmsElectrochimica Acta, 1983
- Surface and bulk processes at oxidized iridium electrodes—I. Monolayer stage and transition to reversible multilayer oxide film behaviourElectrochimica Acta, 1983
- The Suitability of the antimony electrode for pH determinations in mammalian heartJournal of Pharmacological Methods, 1980
- The influence of carbon dioxide, bicarbonate and other buffers on the potential of antimony microelectrodesPflügers Archiv - European Journal of Physiology, 1977
- On the Mixed Potentials Observed in the Iridium-Oxygen-Acid SystemJournal of the Electrochemical Society, 1964
- The Effect of Metal Dissolution on the Rest Potential in the Palladium-Oxygen-Acid SystemJournal of the Electrochemical Society, 1964
- THE ANTIMONY-ANTIMONY TRIOXIDE ELECTRODE AND ITS USE AS A MEASURE OF ACIDITY1Journal of the American Chemical Society, 1928