Distortion of non-invasive cardiac pulse curves. A capillary-damped pick-up and a calibration unit forapex cardiograms and other pulse curves.
Open Access
- 1 September 1977
- Vol. 39 (9) , 995-1005
- https://doi.org/10.1136/hrt.39.9.995
Abstract
A calibration unit and a capillary-damped pick-up, of funnel type, intended for apex cardiograms and other non-invasive cardiac pulse curves are presented. The physical properties of a recording system, suitable for non-invasive cardiac diagnostic techniques, are reviewed. In 4 identical recording units consisting of a funnel with and without capillary damping, connected via air transmission to a transducer, the overamplification in the underdamped systems varied between 2-4 and 4-3, and in the capillary-damped systems between 1-1 and 1-2. The capillary damping gave an approximately linear relation between phase shift and frequency in the high frequency range with a calculated delay of about 4ms. The low frequency time constant in the systems varied between 1-9 and 4-6 s, depending on the amplifier setting used, which for sine curves of frequency 0-5 and 2 Hz means an apparent prematurity for their maxima and minima varying between 53 and 1 ms, this being larger with lower frequency time constants and lower frequency curves.This publication has 9 references indexed in Scilit:
- A Computerized System for Optimal Filtering of Left Ventricular Pressure DataIEEE Transactions on Biomedical Engineering, 1975
- On the clinical value of calibrated displacement apexcardiography.Circulation, 1975
- A combined microphone for simultaneous recording of pulse and heart sounds.1973
- Characteristics of transducers used for recording the apexcardiogram.Journal of Applied Physiology, 1971
- Air Leaks as a Source of Distortion in ApexcardiographyChest, 1970
- Relationship Between Quantitated Precordial Movement and Left Ventricular FunctionCirculation, 1970
- On the physical principles and methodology of mechanocardiography.1969
- Evaluation of the Contractile State of the Human Heart from the First Derivative of the ApexcardiogramCirculation, 1967
- Quantitation of Precordial MovementCirculation, 1967