Measurement of cardioversion/defibrillation thresholds in man by a truncated exponential waveform and an apical patch-superior vena caval spring electrode configuration.
- 1 April 1984
- journal article
- research article
- Published by Wolters Kluwer Health in Circulation
- Vol. 69 (4) , 766-771
- https://doi.org/10.1161/01.cir.69.4.766
Abstract
Defibrillation/cardioversion thresholds were measured in 33 patients undergoing defibrillator implants. Each patient had a 12 cm2 patch placed near the left ventricular apex via a thoracotomy and a 10 cm2 spring lead placed pervenously at the right atrial-superior vena caval junction. Ventricular tachycardia of stable morphology, polymorphic ventricular tachycardia, or ventricular fibrillation was induced four times in each patient and 1, 5, 10, and 25 J truncated exponential shocks with 60% tilt were given in a random sequence. The conversion rate was constant (77%, 86%, 87%, 85%) with increasing energy for ventricular tachycardia but progressively increased for polymorphic ventricular tachycardia and ventricular fibrillation (8%, 33%, 58%, 92%). The ventricular tachycardia acceleration rates for 1, 5, 10, and 25 J were 23%, 14%, 10%, and 15%. Patients not reliably converted with 25 J may require repositioning of leads or two patches. We conclude that for the spring-patch electrodes, increasing energy from 1 to 25 J improves the conversion rate for polymorphic ventricular tachycardia and ventricular fibrillation; the ventricular tachycardia conversion rate is constant. Acceleration of ventricular tachycardia occurs at all energies. Defibrillator implantation requires extensive intraoperative electrophysiologic testing to ensure safe and reliable termination of ventricular tachycardia and fibrillation.This publication has 12 references indexed in Scilit:
- Role of a catheter lead system for transvenous countershock and pacing during electrophysiologic tests: An assessment of the usefulness of catheter shocks for terminating ventricular tachyarrhythmiasThe American Journal of Cardiology, 1983
- The automatic implantable defibrillator: Local ventricular bipolar sensing to detect ventricular tachycardia and fibrillationThe American Journal of Cardiology, 1983
- Clinical evaluation of the internal automatic cardioverter-defibrillator in survivors of sudden cardiac deathThe American Journal of Cardiology, 1983
- Clinical transvenous cardioversion of recurrent life-threatening ventricular tachyarrhythmias: Low energy synchronized cardioversion of ventricular tachycardia and termination of ventricular fibrillation in patients using a catheter electrodeAmerican Heart Journal, 1982
- Clinical treatment of life-threatening ventricular tachyarrhythmias with the automatic implantable defibrillatorAmerican Heart Journal, 1981
- Low energy electrical defibrillation of human hearts during cardiac surgeryThe Journal of Thoracic and Cardiovascular Surgery, 1974
- The electrical dose for ventricular defibrillation with electrodes applied directly to the heartThe Journal of Thoracic and Cardiovascular Surgery, 1974
- Relationship between electrode geometry and effectiveness of ventricular defibrillation in the dog with catheter having one electrode in right ventricle and other electrode in superior vena cava, or external jugular vein, or bothCardiovascular Research, 1973
- Standby Automatic DefibrillatorArchives of internal medicine (1960), 1970