Fluoroscope-generated electromagnetic interference in an external demand pacemaker: report of a case.
- 1 March 1977
- journal article
- research article
- Published by Wolters Kluwer Health in Circulation
- Vol. 55 (3) , 548-550
- https://doi.org/10.1161/01.cir.55.3.548
Abstract
Electromagnetic interference presented as inhibition and resetting of the demand circuitry of a ventricular-inhibited temporary external pacemaker in a 70-year-old man undergoing surgical implantation of a permanent bipolar pacemaker generator and lead. The arrhythmia was found to be due to oversensing of waveforms modulated to simulate myocardial potentials emitted by a faulty fluoroscopy unit used in the vicinity of external temporary transvenous pacing equipment. The documentation of this disruption of pacemaker rhythm reinforces the need for continuous monitoring of patients treated with external demand pacemakers and for the careful maintenance of all electrical equipment.This publication has 12 references indexed in Scilit:
- The Treatment of Acute Traumatic Rupture of the AortaAnnals of Surgery, 1976
- Pacemaker triggering (inhibition) by electric toothbrushThe American Journal of Cardiology, 1976
- Suppression of a ventricular-inhibited bipolar pacemaker by skeletal muscle activityJournal of Electrocardiology, 1974
- Evaluation of normal and abnormal sensing functions of demand pacemakersThe American Journal of Cardiology, 1971
- Electrical interference in non-competitive pacemakersHeart, 1970
- The Influence of Electromagnetic Environment on the Performance of Artificial Cardiac PacemakersThe Annals of Thoracic Surgery, 1968
- Traumatic left ventricular aneurysmThe American Journal of Cardiology, 1967
- CONTUSIONS OF THE HEARTPublished by Wolters Kluwer Health ,1967
- Traumatic rupture of the heart and intrapericardial structuresThe American Journal of Surgery, 1935