Wear-Through of a Modular Polyethylene Liner
- 1 February 2001
- journal article
- case report
- Published by Wolters Kluwer Health in Clinical Orthopaedics and Related Research
- Vol. 383 (383) , 175-182
- https://doi.org/10.1097/00003086-200102000-00019
Abstract
Four patients in whom the S-ROM Total Hip System Polydial polyethylene liner was used illustrate the importance of, and difficulties in, detecting polyethylene wear-through before a complete acetabular revision is necessary. The patients in these cases had full thickness polyethylene wear-through 11 or more years after total hip arthroplasties. Dorr and Livermore wear measurements from radiographs of the patients underestimated the actual wear and did not indicate wear-through. Because excessive head penetration was not evident radiographically with the S-ROM components, the cases showed the necessity of using additional measures to diagnose wear-through, particularly with these components. Repeat radiographs at a higher kilovoltage were used to better visualize the outline of the femoral head, and hip aspiration was used to diagnose metallosis. These cases also showed the importance of yearly followup evaluations, especially 10 or more years after surgery for patients with the S-ROM Polydial locking mechanism.Keywords
This publication has 9 references indexed in Scilit:
- Polyethylene wear, osteolysis and acetabular loosening with an HA-coated hip prosthesisThe Journal of Bone and Joint Surgery. British volume, 1999
- Late failure of the polyethylene liner fixation in an uncemented total hip arthroplastyThe Journal of Arthroplasty, 1999
- The Porous-Coated Anatomic Total Hip Prosthesis, Inserted without Cement. A Prospective Study with a Minimum Ten Years of Follow-up*Journal of Bone and Joint Surgery, 1999
- Treatment of Pelvic Osteolysis Associated with a Stable Acetabular Component Inserted without Cement as Part of a Total Hip Replacement*Journal of Bone and Joint Surgery, 1997
- Primary Total Hip Arthroplasty with a Porous-Coated Acetabular Component. Seven-to-Ten-Year Results*Journal of Bone and Joint Surgery, 1997
- Comparative results of a distal modular sleeve, circumferential coating, and stiffness relief using the Anatomic Porous Replacement IIThe Journal of Arthroplasty, 1996
- Catastrophic failure of the polyethylene liner of uncemented acetabular componentsThe Journal of Bone and Joint Surgery. British volume, 1994
- Severe polyethylene wear in total hip arthroplastyThe Journal of Arthroplasty, 1994
- Ectopic Ossification Following Total Hip ReplacementJournal of Bone and Joint Surgery, 1973