Accuracy, reproducibility, and variability of hand-held dynamometry in motor neuron disease.
Open Access
- 1 March 1994
- journal article
- research article
- Published by BMJ in Journal of Neurology, Neurosurgery & Psychiatry
- Vol. 57 (3) , 326-332
- https://doi.org/10.1136/jnnp.57.3.326
Abstract
A spring-loaded device that "breaks" at preset forces was used to assess readings obtained by hand-held dynamometry by three raters with varying experience in the method. Overall accuracy (3%), but not reproducibility or variability, was improved by greater experience. Readings obtained jointly by three raters had 53% greater variability than those obtained by a single rater. Nine muscle groups in 19 patients with motor neuron disease were assessed at 10 sessions (three replications per session) over six days by the experienced rater. Muscle force was expressed relative to that of 22 matched normal controls. The reproducibility was good with a mean % difference of 13.2 and repeatability coefficient of 2.17 kg-force for readings six days apart; the overall correlation coefficient was 0.98. The mean coefficient of variation (CV) of 10 readings was 9.9%. The poorer reproducibility and greater variability seen in clinically weaker muscles may account for differences in patients with bulbar palsy and classical amyotrophic lateral sclerosis; the degree of spasticity had no effect. The rater was estimated to contribute 37% of the total variability when testing patients. The use of a composite score by combining normalised dynamometry readings of eight limb muscles improved mean % difference to 6.7 and mean CV to 5.8%. The reproducibility and variability of hand-held dynamometry readings obtained by a single rater compare well with those of fixed devices. Readings from single raters, irrespective of experience, have similar reproducibility and variability. If, however, multiple raters are used in longitudinal assessments of individual patients, as occurs in clinical trials, the variability of their combined readings should be estimated when calculating the same size required.Keywords
This publication has 21 references indexed in Scilit:
- Hand-held myometry: reference values.Journal of Neurology, Neurosurgery & Psychiatry, 1991
- Controlled acute trial of a thyrotrophin releasing hormone analogue (RX77368) in motor neuron disease.Journal of Neurology, Neurosurgery & Psychiatry, 1987
- STRENGTH TESTING WITH A PORTABLE DYNAMOMETER - RELIABILITY FOR UPPER AND LOWER-EXTREMITIES1987
- Observations on the Clinical Assessment of Patients with Motor Neuron Disease: Experience with a TRH AnalogueNeurologic Clinics, 1987
- Quantitative motor assessment in amyotrophic lateral sclerosisNeurology, 1986
- Measuring muscle strengthZeitschrift für Neurologie, 1984
- The measurement of muscle strength in patients with peripheral neuromuscular disorders.Journal of Neurology, Neurosurgery & Psychiatry, 1983
- Amyotrophic Lateral SclerosisArchives of Neurology, 1979
- Measurements of muscle strength and performance in children with normal and diseased muscle.Archives of Disease in Childhood, 1976
- HAND-HELD DYNAMOMETER FOR EVALUATING VOLUNTARY-MUSCLE FUNCTIONThe Lancet, 1974