Efficiency of a Bismuth-Germanate Scintillator: Comparison of Monte Carlo Calculations with Measurements
- 1 January 1984
- journal article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Nuclear Science
- Vol. 31 (1) , 390-395
- https://doi.org/10.1109/tns.1984.4333284
Abstract
Monte Carlo calculations of a bismuth-germanate scintillator's efficiency agree closely with experimental measurements. For this comparison, we studied the absolute gamma-ray photopeak efficiency of a scintillator (7.62 cm long by 7.62 cm in diameter) at several gamma-ray energies from 166 to 2615 keV at distances from 30.5 to 152.4 cm. Computer calculations were done in a three-dimensional cylindrical geometry with the Monte Carlo coupled photon-electron code CYLTRAN. For the experiment we measured 11 sources with simple spectra and precisely known strengths. The average deviation between the calculations and the measurements is 3%. Our calculated results also closely agree with recently published calculated results.Keywords
This publication has 3 references indexed in Scilit:
- Unfolding bismuth-germanate pulse-height distributions to determine gamma-ray flux spectra and dose ratesNuclear Instruments and Methods in Physics Research, 1984
- Bismuth Germanate as a High-Energy Gamma-Ray DetectorIEEE Transactions on Nuclear Science, 1983