Abstract
The natural variability in forced expiratory volume in one second (FEV1) over 20 minutes was determined in 54 fit hospital employees and 13 patients with restrictive lung disorders. Initial FEV1 ranged from 1.1 to 6.3 1 BTPS. Variability when expressed as absolute change was similar at all levels of FEV1, so that, when expressed as percentage change, variability decreased with increasing FEV1. Smoking habits did not appear to affect variability but activity before the test did. On the basis of these results an absolute change in FEV1 of 190 ml would be necessary for 95% confidence that the change in FEV1 occurred other than by chance in any one individual. This suggests that the absolute change in FEV1 might be a more reliable criterion than percentage change when distinguishing between natural variability and a response to inhalation of bronchodilators.This publication has 8 references indexed in Scilit:
- Estimation of the FEV.Thorax, 1983
- How many blows really make an FEV1, FVC, or PEFR?Thorax, 1983
- ABSOLUTE FEV1 VALUESThe Lancet, 1981
- RANDOM ERROR WITH THE FEV1 = CASE FOR ABSOLUTE VALUESThe Lancet, 1981
- WITHIN-SUBJECT VARIABILITY AND PERCENT CHANGE FOR SIGNIFICANCE OF SPIROMETRY IN NORMAL SUBJECTS AND IN PATIENTS WITH CYSTIC-FIBROSISPublished by Elsevier ,1980
- EPIDEMIOLOGY STANDARDIZATION PROJECT1978
- Variability of routine pulmonary function tests.Thorax, 1975
- Assessment of ventilatory abnormality in the asymptomatic subject: an exercise in futility.Thorax, 1966