Diffusion-weighted Imaging with Navigated Interleaved Echo-planar Imaging and a Conventional Gradient System

Abstract
PURPOSE: To demonstrate the technical feasibility and precision of a navigated diffusion-weighted (DW) MR imaging method with interleaved echo-planar imaging and test its diagnostic sensitivity for detection of ischemic stroke. MATERIALS AND METHODS: Apparent diffusion coefficient (ADC) measurements were performed in phantoms, and six healthy adult volunteers were examined to determine intrasubject (precision) and intersubject (reference range) variations in absolute ADC and relative ADC (rADC) measurements. DW imaging maps and lesion rADC values were also obtained in 34 consecutive stroke patients to evaluate the sensitivity and reliability of DW-interleaved echo-planar imaging for detection of ischemic brain damage. RESULTS: Phantom and volunteer ADC values were in excellent agreement with published data. The intrasubject variation of rADC was 6.2%. The ADC precision ranged from 6.5% in the subcortical white matter in the frontal lobe to 12.9% in the head of the caudate nucleus. Interleaved echo-planar ...