High-performance liquid affinity chromatography: rapid immunoanalysis of transferrin in serum.
Open Access
- 1 October 1988
- journal article
- research article
- Published by Oxford University Press (OUP) in Clinical Chemistry
- Vol. 34 (10) , 2039-2043
- https://doi.org/10.1093/clinchem/34.10.2039
Abstract
We describe a new method for quantitatively measuring substances of clinical interest by high-performance liquid affinity chromatography (HPLAC). As a model system we selected analysis for transferrin in human serum with immobilized antibodies in a high-performance liquid chromatographic system. SelectiSpher-10 Activated Tresyl columns (5 or 10 x 0.5 cm) were used for in situ coupling of polyclonal antibodies to transferrin. The amount of transferrin eluted was determined by integrating the eluted peak at 280 nm. The whole analytical procedure--including injection of sample, washing, elution, and analysis of data--takes only 7 min. We characterized the HPLAC system for analysis of transferrin in several ways: intra-assay CV approximately 3%; inter-assay CV 2-9%; linear response up to 1 mg/mL column volume; detection limit approximately 3 micrograms; analytical recovery 98% +/- 2%; purity of eluted sample greater than 95% (SDS-PAGE). The HPLAC method was compared with "rocket" immunoelectrophoresis, a commonly used method of analysis for transferrin, and there was excellent correlation between the two methods (r = 0.96, n = 60). Benefits of this HPLAC technique include high precision, rapid analysis, and simplified sample handling.This publication has 0 references indexed in Scilit: