Study of Pacemaker and Implantable Cardioverter Defibrillator Triggering by Electronic Article Surveillance Devices (SPICED TEAS)
- 1 October 1998
- journal article
- research article
- Published by Wiley in Pacing and Clinical Electrophysiology
- Vol. 21 (10) , 1847-1861
- https://doi.org/10.1111/j.1540-8159.1998.tb00002.x
Abstract
The magnetic fields emitted by electronic article surveillance (EAS) systems (shoplifting gates) are a source of interference for implanted medical devices. In the Study of Pacemaker and Implantable Cardioverter Defibrillator Triggering by Electronic Article Surveillance Devices (SPICED TEAS), 25 adult volunteers with ICDs and 50 with pacemakers were exposed to the fields of six different EAS systems. These EAS systems used three modes of operation: magnetic audio frequency, swept radiofrequency, and acoustomagnetic technology. No ICD exhibited interference mimicking sensing of tachyarrhythmias with any EAS system. Pacemakers interacted variably, depending on the type of EAS system. Swept radiofrequency systems produced no interaction with any implanted medical device. One magnetic audio frequency system interacted with 2 of 50 pacemakers. The acoustomagnetic system interacted with 48 of 50 pacemakers. Interactions included asynchronous pacing, atrial oversensing (producing "EAS induced tachycardia" in the ventricle), ventricular oversensing (with pacemaker inhibition), and paced beats resulting from the direct induction of current in the pacemaker ("EAS induced pacing"). These interactions produced symptoms in some patients (palpitations, presyncope) only while patients were in the EAS field. No pacemaker was reprogrammed. We conclude that high energy, pulsed low frequency EAS systems such as acoustomagnetic systems interfere with most pacemakers. Pacemaker patients should be advised to minimize exposure to the fields of such systems to prevent the possibility of serious clinical events.Keywords
This publication has 7 references indexed in Scilit:
- Interaction Between Electronic Article Surveillance Systems and Implantable Defibrillators: Insights from a Fourth Generation ICDPacing and Clinical Electrophysiology, 1997
- Environmental Electromagnetic Interference from Electronic Article Surveillance Devices: Interactions with an ICDPacing and Clinical Electrophysiology, 1995
- The Effects of Electronic Article Surveillance Systems on Permanent Cardiac Pacemakers: An In Vitro StudyPacing and Clinical Electrophysiology, 1994
- Interference From a Hand Held Radiofrequency Remote Control Causing Discharge of an Implantahle DefibrillatorPacing and Clinical Electrophysiology, 1993
- Electronic Article Surveillance: A Possible Danger for Pacemaker PatientsPacing and Clinical Electrophysiology, 1993
- Interference in PacemakersPacing and Clinical Electrophysiology, 1984
- The Influence of Electromagnetic Environment on the Performance of Artificial Cardiac PacemakersThe Annals of Thoracic Surgery, 1968