Hughes External Fixator in Treatment of Tibial Fractures
- 1 October 1985
- journal article
- research article
- Published by SAGE Publications in Journal of the Royal Society of Medicine
- Vol. 78 (10) , 830-837
- https://doi.org/10.1177/014107688507801008
Abstract
The results of a prospective trial of the use of the Hughes unilateral external fixator in the management of 48 tibial diaphyseal fractures are presented. Good results were obtained in grade II and III fractures but not in closed and grade I fractures. The results and complication rates were comparable with those of other more complex external fixators. Good results were found to be dependent on the adequacy of the initial reduction and the duration of external fixation, but independent of alterations in pin angle, length and location as well as fixator location.This publication has 28 references indexed in Scilit:
- The treatment of fractures with a dynamic axial fixator.1984
- Comparison of Mechanical Performance in Four Types of External FixatorsClinical Orthopaedics and Related Research, 1983
- External fixation of severe open tibial fractures with the Hoffmann frame.1983
- Tibial Fractures Treated with Hoffmann's External Fixation:A Comparative Analysis of Hoffmann Bilateral Frames and the Vidal-Adrey Double Frame ModificationActa Orthopaedica, 1982
- Which external fixation device?Injury, 1981
- How to increase the stability of external fixation unitsArchives of orthopaedic and trauma surgery, 1980
- Analysis of Single-Sided External Fracture FixationEngineering in Medicine, 1979
- Cavus deformity of the foot after fracture of the tibial shaftJournal of Bone and Joint Surgery, 1975
- A Functional Below-the-Knee Cast for Tibial FracturesJournal of Bone and Joint Surgery, 1967
- Rigidity of Immobilization of Transverse FracturesActa Orthopaedica, 1962