Relation of mean pixel intensity to concentration of sonicated albumin microspheres: Effects of ultrasound system settings
Open Access
- 1 November 1991
- journal article
- research article
- Published by Wiley in Clinical Cardiology
- Vol. 14 (S5) , V-23-23
- https://doi.org/10.1002/clc.4960141706
Abstract
Myocardial contrast echocardiography (MCE) is a recently developed technique to assess myocardial perfusion. Previous studies have shown that MCE frequently underestimates coronary reserve as compared with other methods. Previous reports have suggested that some of the limitations of MCE are due to the nonlinear processing algorithms of the ultrasound systems. Therefore we designed an in vitro model to study the effects of various processing algorithms and levels of acoustic power (AP) on the relation between concentration of echocontrast and mean pixel intensities. A beaker containing a commercially produced, diluted solution of sonicated albumin microspheres was imaged with a commercial Hewlett Packard ultrasound system, and mean pixel intensities were determined with an off-line computer system. The solution was imaged at a wide range of APs (40–10 dB) and time gain compensation (40–60 dB). Concentration versus intensity curves were generated using all possible combinations of compression and processing algorithms available on the ultrasound system used. Contrast effect diminished rapidly when exposed to high AP. Lower powers had less effect on contrast intensity duration, with no effect seen below 20 dB. Changes in time gain compensation did not affect contrast intensities. Regardless of the processing algorithm examined, the relation between concentration of echocontrast and mean pixel intensities was near linear in the first two to three concentrations; thereafter, the curves flattened during increasing concentrations of contrast. At higher concentrations, acoustic shadowing produced a decrease in pixel intensities. Thus, in this in vitro model, measurements of pixel intensities offer the opportunity of quantitating sonicated albumin microsphere concentration, but only at the lower range (6.25 × 10−5/ml sonicated albumin microspheres/ml of solution) of the contrast concentration. These results could have important clinical implications.Keywords
This publication has 20 references indexed in Scilit:
- Assessment of regional myocardial perfusion by contrast echocardiography. II. Detection of changes in transmural and subendocardial perfusion during dipyridamole-induced hyperemia in a model of critical coronary stenosisJournal of the American College of Cardiology, 1989
- Quantitative assessment of the immediate results of coronary angioplasty by myocardial contrast echocardiographyJournal of the American College of Cardiology, 1989
- Assessment of regional myocardial blood flow with myocardial contrast two-dimensional echocardiographyJournal of the American College of Cardiology, 1989
- Myocardial contrast echocardiography in humans. II. Assessment of coronary blood flow reserveJournal of the American College of Cardiology, 1988
- Assessment of myocardial perfusion in humans by contrast echocardiography. I. Evaluation of regional coronary reserve by peak contrast intensityJournal of the American College of Cardiology, 1988
- Contrast echocardiographic estimation of regional myocardial blood flow after acute coronary occlusion.Circulation, 1985
- Contrast echocardiography in acute myocardial ischemia: I. In vivo determination of total left ventricular “area at risk”Journal of the American College of Cardiology, 1984
- Hydrogen peroxide contrast--enhanced two-dimensional echocardiography: real-time in vivo delineation of regional myocardial perfusion.Circulation, 1983
- Assessment of location and size of myocardial infarction with contrast-enhanced echocardiographyJournal of the American College of Cardiology, 1983
- Assessment of myocardial perfusion abnormalities with contrast-enhanced two-dimensional echocardiography.Circulation, 1982