Human factors engineering and patient safety
- 1 December 2002
- journal article
- Published by BMJ in Quality and Safety in Health Care
- Vol. 11 (4) , 352-354
- https://doi.org/10.1136/qhc.11.4.352
Abstract
The case study and analyses presented here illustrate the crucial role of human factors engineering (HFE) in patient safety. HFE is a framework for efficient and constructive thinking which includes methods and tools to help healthcare teams perform patient safety analyses, such as root cause analyses. The literature on HFE over several decades contains theories and applied studies to help to solve difficult patient safety problems and design issues. A case study is presented which illustrates the vulnerabilities of human factors design in a transport monitor. The subsequent analysis highlights how to move beyond the more obvious contributing factors like training to design problems and the establishment of informal norms. General advice is offered to address these issues and design issues specific to this case are discussed.Keywords
This publication has 8 references indexed in Scilit:
- Safe but SoundJAMA, 2002
- The Wrong PatientAnnals of Internal Medicine, 2002
- How to investigate and analyse clinical incidents: Clinical Risk Unit and Association of Litigation and Risk Management protocolBMJ, 2000
- Human factors in the health care facility.1998
- Human error and human factors engineering in health care.1998
- Human factors and the FDA's goals: improved medical device design.1996
- A Monitor with a Mind of Its OwnAnesthesiology, 1995
- What are we really monitoring?1994