Evaluating data acquisition and smoothing functions of currently available videokeratoscopes
- 1 May 1996
- journal article
- research article
- Published by Wolters Kluwer Health in Journal of Cataract & Refractive Surgery
- Vol. 22 (4) , 421-426
- https://doi.org/10.1016/s0886-3350(96)80036-8
Abstract
Purpose: To compare the accuracy of computerized videokeratography systems using identical, calibrated test objects. Setting: Lions Eye Institute, Albany, New York. Methods: We evaluated the accuracy and smoothing of raw data acquisition (axial solution) of seven commercially available videokeratoscopes: Alcon EyeMap, Computed Anatomy TMS, EyeSys CAS, Humphrey MasterVue, Topcon CM-1000, Optikon Keratron, and TechnoMed C-Scan. We used six calibrated test objects to simulate clinical settings: spherical, spherocylindrical, simulated myopic ablation, hyperopic ablation, and a simulated central island. Results: None of the systems accurately imaged all objects. Although all systems imaged spherical objects with reasonable accuracy, errors greater than 4.0 diopters (D) frequently occurred in the central 6.0 mm optical zone (maximum error 10.0 D). Sources of error included excessive raw data smoothing, inability to read large transitions, loss of accuracy in the periphery, and poor central coverage. Conclusion: The clinician should be aware of the potential limitations of corneal topography when making clinical decisions.Keywords
This publication has 3 references indexed in Scilit:
- Characterization of the Inherent Error in a Spherically-Biased Corneal Topography System in Mapping a Radially Aspheric Surface/Comment: Robert B. Mandell, OD, PhD Berkeley, Calif/Comment: Raymond A. Applegate, OD, PhD San Antonio, Tex/Response: Cynthia Roberts, PhDJournal of Refractive Surgery, 1994
- Comparison of the accuracy and reproducibility of the keratometer and the EyeSys Corneal Analysis System Model IJournal of Cataract & Refractive Surgery, 1992
- Accuracy and Precision of Keratometry, Photokeratoscopy, and Corneal Modeling on Calibrated Steel BallsArchives of Ophthalmology (1950), 1989