Distortion of cascade impactor measured size distribution due to bounce and blow-off
- 4 June 1980
- journal article
- research article
- Published by Taylor & Francis in Aihaj Journal
- Vol. 41 (6) , 410-419
- https://doi.org/10.1080/15298668091424951
Abstract
Accurate measurement of particle size distribution by a cascade impactor requires that particles remain on the collection surfaces after impaction. Recent studies show that, when a solid aerosol is sampled, significant rebound and re-entrainment of particles occur in cascade impactors. A mathematical model of a cascade impactor including collection efficiency functions for solid aerosols on various impaction surfaces was developed and used to operate upon a series of log-normal size distributions. The impactor modeled was a six stage Andersen operated at 28.3 L/min. When bare metal or glass impaction surfaces are used, the impactor indicates non-log-normal distributions and the estimated distribution parameters are in error by as much as a factor of 7. When glass fiber impaction surfaces are used, although less serious errors are indicated, the deviation from log-normality is significant for aerosols with MMD larger than 3 micro;m. When oil covered impaction stages are used the indicated distribution parameters are acceptably close to the input parameters. The accuracy of the model was verified experimentally.This publication has 2 references indexed in Scilit:
- Non-ideal collection characteristics of inertial impactors—II. Cascade impactorsJournal of Aerosol Science, 1978
- Non-ideal collection characteristics of inertial impactors—I. Single-stage impactors and solid particlesJournal of Aerosol Science, 1978