Cardiac CINE MR imaging with a 32‐channel cardiac coil and parallel imaging: Impact of acceleration factors on image quality and volumetric accuracy
Open Access
- 22 December 2005
- journal article
- research article
- Published by Wiley in Journal of Magnetic Resonance Imaging
- Vol. 23 (2) , 222-227
- https://doi.org/10.1002/jmri.20484
Abstract
Purpose To assess the impact of parallel imaging algorithms on image quality and volumetric accuracy of CINE magnetic resonance imaging (MRI) with high temporal and spatial resolution using a new 32‐channel dedicated cardiac phased array coil. Materials and Methods Fourteen individuals underwent steady‐state free precession (SSFP) CINE MRI using a 32‐element phased‐array coil and parallel imaging acceleration using spatiotemporal sensitivity encoding (TSENSE). Acquisition acceleration ranged from R = 2 to 7. In conjunction with data extracted from phantom measurements, contrast‐to‐noise ratio (CNR) performance was evaluated for each acceleration factor and subjective image quality was evaluated by two independent readers. In addition, volumetric assessment was performed for each acceleration factor based on a single breath‐hold multi‐slice data acquisition. Results were compared to nonTSENSE measurements. Results CNR for non‐accelerated CINE (R = 1) was 45.7 ± 12.8 and showed a constant decrease with increase in acceleration of 51% at R = 4 and 86% at R = 7. CNR losses accompanied reductions in subjective image quality. Volumetric evaluation was accurate for R ≤ 4, with significant underestimation of ejection fraction (EF) at higher accelerations. Conclusion This study shows that one‐dimensional acceleration factors up to R = 4 allow accurate SSFP CINE MRI even though CNR is significantly reduced. This allows for a marked reduction in scan time and allows for multi‐slice CINE imaging with high spatial and temporal resolution. J. Magn. Reson. Imaging 2006.Keywords
This publication has 19 references indexed in Scilit:
- Highly parallel volumetric imaging with a 32‐element RF coil arrayMagnetic Resonance in Medicine, 2004
- Single breath-hold real-time cine MR imaging: improved temporal resolution using generalized autocalibrating partially parallel acquisition (GRAPPA) algorithmEuropean Radiology, 2003
- Breath-hold FLASH and FISP Cardiovascular MR Imaging: Left Ventricular Volume Differences and ReproducibilityRadiology, 2002
- MR Imaging of the Heart with Cine True Fast Imaging with Steady-State Precession: Influence of Spatial and Temporal Resolutions on Left Ventricular Functional ParametersRadiology, 2002
- Cardiac Function: MR Evaluation in One Breath Hold with Real-time True Fast Imaging with Steady-State PrecessionRadiology, 2002
- Assessment of Ventricular Function with Single Breath-Hold Real-Time Steady-State Free Precession Cine MR ImagingAmerican Journal of Roentgenology, 2002
- Functional cardiac MR imaging with steady‐state free precession (SSFP) significantly improves endocardial border delineation without contrast agentsJournal of Magnetic Resonance Imaging, 2001
- MR Evaluation of Ventricular Function: True Fast Imaging with Steady-State Precession versus Fast Low-Angle Shot Cine MR Imaging: Feasibility StudyRadiology, 2001
- SENSE: Sensitivity encoding for fast MRIMagnetic Resonance in Medicine, 1999
- Simultaneous acquisition of spatial harmonics (SMASH): Fast imaging with radiofrequency coil arraysMagnetic Resonance in Medicine, 1997