Investigation of matrix effects for Pb isotope ratio measurements by multiple collector ICP-MS: verification and application of optimized analytical protocols
- 25 October 2000
- journal article
- research article
- Published by Royal Society of Chemistry (RSC) in Journal of Analytical Atomic Spectrometry
- Vol. 15 (11) , 1451-1460
- https://doi.org/10.1039/b005262k
Abstract
Previous studies have demonstrated the potential of multiple-collector inductively coupled plasma mass spectrometry (MC-ICP-MS) for precise Pb isotopic measurements using admixed Tl for "external normalization" of instrumental mass discrimination. The Micromass IsoProbe is a new, single-focusing MC-ICP-MS instrument that does not employ an electrostatic sector for energy-focusing. Instead, a hexapole collision cell is used to thermalize the ion beam. This study presents the first in-depth investigation of the application of the IsoProbe to Pb isotope ratio measurements. With a Tl-based mass bias correction, multiple analyses of mixed Pb–Tl standard solutions typically display reproducibilities (±2σ) of about 50 ppm for 207Pb∶206Pb, 100 ppm for 208Pb∶206Pb and 206Pb∶204Pb, 150 ppm for 207Pb∶204Pb and 175 ppm for 208Pb∶204Pb. Using an empirically optimized exponential law normalization, the Pb isotope data show excellent agreement with reference values obtained by thermal ionization mass spectrometry. The cross-calibration between NIST SRM 981 and 982 demonstrates that accurate results can be obtained with this technique over a wide range of Pb isotopic compositions. The Tl-corrected Pb data obtained on the IsoProbe, however, display correlations with the Pb∶Tl ratio of the analyzed solutions. The addition of large amounts of Tl to the samples is unfavorable for the accurate measurement of the low-intensity 204Pb ion beam. Analytical artifacts may also be generated if high concentrations of concomitant elements are present in the sample solutions. Matrix effects and instrumental memory, however, can be readily overcome by the adoption of appropriate analytical protocols and it is demonstrated that sample measurements can achieve the same levels of precision and accuracy as are routinely obtained for analyses of pure standard solutions.Keywords
This publication has 16 references indexed in Scilit:
- High-precision analysis of Pb isotope ratios by multi-collector ICP-MSChemical Geology, 2000
- Inter-laboratory and other errors in Pb isotope analyses investigated using a 207Pb–204Pb double spikeChemical Geology, 2000
- Precise analysis of copper and zinc isotopic compositions by plasma-source mass spectrometryChemical Geology, 1999
- A new variable dispersion double-focusing plasma mass spectrometer with performance illustrated for Pb isotopesInternational Journal of Mass Spectrometry, 1998
- Practical Application of Lead Triple Spiking for Correction of Instrumental Mass DiscriminationMineralogical Magazine, 1998
- Inductively coupled plasma magnetic sector mass spectrometry method for stable lead isotope tracer studiesJournal of Analytical Atomic Spectrometry, 1998
- Lead isotopic analyses of NIST Standard Reference Materials using multiple collector inductively coupled plasma mass spectrometry coupled with a modified external correction method for mass discrimination effectThe Analyst, 1996
- Verification of a correction procedure for measurement of lead isotope ratios by inductively coupled plasma mass spectrometryJournal of Analytical Atomic Spectrometry, 1991
- Revised isotopic composition of terrestrial mercuryInternational Journal of Mass Spectrometry and Ion Processes, 1989
- Determination of lead isotope ratios by inductively coupled plasma-mass spectrometry (ICP-MS)Spectrochimica Acta Part B: Atomic Spectroscopy, 1987