Attack and release times of automatic-gain-control hearing aids
- 1 September 1977
- journal article
- research article
- Published by Acoustical Society of America (ASA) in The Journal of the Acoustical Society of America
- Vol. 62 (3) , 784-786
- https://doi.org/10.1121/1.381554
Abstract
The attack and release times were measured on each of three samples of 27 different models of compression-type hearing aids using the test conditions specified in ANSI Standard S3-22 (1976). The results indicate that almost all of the hearing aids tested have attack times less than or equal to 10 ms. A little more than half of the hearing aids had release times of 50 ms or less. The range of the attack times varied from 1 to 23 ms and the release times from 5–600 ms.Keywords