Abstract
In this method for detection and quantification of volatile alcohols by capillary gas chromatography, the serum sample is deproteinized, then directly injected into the gas chromatograph with 1-propanol as the internal standard. The capillary column is a 30-m bonded methylsilicone-coated, fused-silica column. With helium as the carrier gas, the injector inlet is set at a split ratio of 1/30 and the average linear velocity in the column is 25 cm/s. Injector and flame-ionization detector temperatures are 280 degrees C, oven temperature 35 degrees C. Chromatography time is less than 3 min.