An Edge Detection Algorithm for Attenuation Correction in Emission CT
- 1 January 1987
- journal article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Nuclear Science
- Vol. 34 (1) , 309-312
- https://doi.org/10.1109/tns.1987.4337354
Abstract
A new body edge detection algorithm is presented for attenuation compensation based on calculation in emission computed tomography. Based on the differential geometry, the edge contour on the image plane can be calculated directly from the edge detected on the projection data. A new algorithm is; (1) detect edge contour on projection data, (2) smooth noisy contour by retaining lower order Fourier coefficients and (3) transform it to the edge contour on the image. Manipulation of two dimensional array is only needed in step (1) and the rest can be done by manipulation of one dimensional array. Robustness of the algorithm was tested on the threshold level and number of Fourier coefficients. The algorithm was applied to positron emission computed tomography. It can be applied to single photon emission computed tomography as well.Keywords
This publication has 12 references indexed in Scilit:
- Scatter Compensation Techniques for SPECTIEEE Transactions on Nuclear Science, 1985
- Software correction of scatter coincidence in positron CTEuropean Journal of Nuclear Medicine and Molecular Imaging, 1984
- A Monte Carlo Simulation of Compton Scattering in Positron Emission TomographyJournal of Computer Assisted Tomography, 1983
- Correction for Scattered Radiation in a Ring Detector Positron Camera by Integral Transformation of the ProjectionsJournal of Computer Assisted Tomography, 1983
- Determination of Object Contour from Projections for Attenuation Correction in Cranial Positron Emission TomographyJournal of Computer Assisted Tomography, 1982
- A BOUNDARY METHOD FOR ATTENUATION CORRECTION IN POSITRON COMPUTED TOMOGRAPHYJournal of Computer Assisted Tomography, 1981
- Noise Identification and Removal in Positron Imaging SystemsIEEE Transactions on Nuclear Science, 1981
- Positologica: A Positron ECT Device with a Continuously Rotating Detector RingIEEE Transactions on Nuclear Science, 1980
- Corrections for Attenuation, Scattered Radiation, and Random Coincidences in a Ring Detector Positron Emission Transaxial TomographIEEE Transactions on Nuclear Science, 1980
- Positome II: A High Efficiency Positron Imaging Device for Dynamic Brain StudiesIEEE Transactions on Nuclear Science, 1979