Plasma separation using a hollow fiber membrane device

Abstract
A disposable hollow fiber device was evaluated by collecting approximately 550 ml of normal donor plasma (n=43) and by performing sham (n=10) and therapeutic (n=12) plasma exchanges. Blood was processed at 70 ml per min, and plasma flux averaged 23 (collection) and 25 (exchange) ml per min (mean separation efficiencies of 52 and 60%, respectively). The procedures were tolerated well by all donors and patients. The plasma hemoglobin concentration in separated plasma averaged 1 mg per dl, and cell contamination was negligible (mean of 1, 3, and 6 RBCs, platelets and WBCs/μl, respectively). There was no evidence of in vivo classical or alternative pathway complement activation as assessed by total hemolytic complement generation (CH50), alternative pathway hemolytic activity (AP50), C3 conversion, or C5 activation, nor were unexpected changes seen in the results of laboratory tests performed after the procedure. Sieving coefficients during sham plasma exchange averaged as follows: albumin, 1.03; IgM,1.0, IgG, 1.0; IgA, 0.98; factor V, 1.07; factor VII, 0.89; factor VIII, 1.05; and factor IX, 1.19. The device appears to be useful for separation of cell‐free plasma from blood during therapeutic plasma exchange procedures.