Coulter counter model S and model S-plus measurements of mean erythrocyte volume (MCV) are influenced by the mean erythrocyte haemoglobin concentration (MCHC)
- 1 January 1981
- journal article
- research article
- Published by Taylor & Francis in Scandinavian Journal of Clinical and Laboratory Investigation
- Vol. 41 (8) , 717-721
- https://doi.org/10.3109/00365518109090520
Abstract
We found differences between the mean erythrocyte volume (MCV) when measured by the Coulter counter model S-plus and when calculated from the erythrocyte volume fraction obtained by centrifugation and corrected for trapped plasma. The difference was positively correlated to the mean erythrocyte haemoglobin concentration (MCHC) so that MCV by the Coulter S+ is too low at low MCHC and too high at high MCHC. Similar although smaller differences were observed with the Coulter S. The erroneous MCV gives errors in the calculated erythrocyte volume fraction and MCHC. Only in the case of MCHC does the error have any practical significance; too high values are obtained at low concentrations of haemoglobin in the erythrocytes and too low at high concentrations. Hence, MCHC measured by the Coulter S+ contains no useful information.Keywords
This publication has 6 references indexed in Scilit:
- The Coulter Counter Model S Plus- the shape of things to comeClinical and Laboratory Haematology, 2008
- Applications of statistics in clinical chemistry: A critical evaluation of regression linesClinica Chimica Acta; International Journal of Clinical Chemistry, 1975
- Re‐assessment of the Reliability of the HaematocritBritish Journal of Haematology, 1972
- Epitaph for the M.C.H.C.BMJ, 1971
- Elimination of Error in Hematocrit Produced by Excessive EDTAAmerican Journal of Clinical Pathology, 1969
- Electrical Counting and Sizing of Mammalian Cells in SuspensionBiophysical Journal, 1965