Design and performance of a multi-pinhole collimation device for small animal imaging with clinical SPECT and SPECT–CT scanners
- 17 July 2008
- journal article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 53 (15) , 4185-4201
- https://doi.org/10.1088/0031-9155/53/15/012
Abstract
A multi-pinhole collimation device is developed that uses the gamma camera detectors of a clinical SPECT or SPECT-CT scanner to produce high-resolution SPECT images. The device consists of a rotating cylindrical collimator having 22 tungsten pinholes with 0.9 mm diameter apertures and an animal bed inside the collimator that moves linearly to provide helical or ordered-subsets axial sampling. CT images also may be acquired on a SPECT-CT scanner for purposes of image co-registration and SPECT attenuation correction. The device is placed on the patient table of the scanner without attaching to the detectors or scanner gantry. The system geometry is calibrated in-place from point source data and is then used during image reconstruction. The SPECT imaging performance of the device is evaluated with test phantom scans. Spatial resolution from reconstructed point source images is measured to be 0.6 mm full width at half maximum or better. Micro-Derenzo phantom images demonstrate the ability to resolve 0.7 mm diameter rod patterns. The axial slabs of a Micro-Defrise phantom are visualized well. Collimator efficiency exceeds 0.05% at the center of the field of view, and images of a uniform phantom show acceptable uniformity and minimal artifact. The overall simplicity and relatively good imaging performance of the device make it an interesting low-cost alternative to dedicated small animal scanners.Keywords
This publication has 27 references indexed in Scilit:
- Small-Animal Preclinical Nuclear Medicine Instrumentation and MethodologySeminars in Nuclear Medicine, 2008
- Image reconstructionPhysics in Medicine & Biology, 2006
- Small animal SPECT and its place in the matrix of molecular imaging technologiesPhysics in Medicine & Biology, 2005
- Small-Animal X-ray Dose from Micro-CTMolecular Imaging, 2004
- Fundamental image quality limits for microcomputed tomography in small animalsMedical Physics, 2003
- Helical pinhole SPECT for small-animal imaging: a method for addressing sampling completenessIEEE Transactions on Nuclear Science, 2003
- High-resolution SPECT using multipinhole collimationIEEE Transactions on Nuclear Science, 2003
- Pinhole collimation for ultra-high-resolution, small-field-of-view SPECTPhysics in Medicine & Biology, 1994
- Accelerated image reconstruction using ordered subsets of projection dataIEEE Transactions on Medical Imaging, 1994
- Coded aperture imaging with uniformly redundant arraysApplied Optics, 1978