Neutron Response of a Polyethylene-Moderated Superheated Drop Detector
- 1 March 1992
- journal article
- Published by Wolters Kluwer Health in Health Physics
- Vol. 62 (3) , 215-222
- https://doi.org/10.1097/00004032-199203000-00001
Abstract
Monte Carlo neutron transport calculations were performed to obtain the neutron response functions of a superheated drop detector (SDD-100) that is surrounded by polyethylene of various thicknesses. The calculational results suggest that a SDD-100 surrounded by 2.54 cm of polyethylene gives the best dose-equivalent response in a free neutron field. Results were confirmed by a sequence of experiments using various neutron sources. This polyethylene-moderated SDD-100 is ideal for use as an area neutron dose monitor.Keywords