The Monostrut Strength
- 1 January 1985
- journal article
- research article
- Published by Taylor & Francis in Scandinavian Journal of Thoracic and Cardiovascular Surgery
- Vol. 19 (1) , 13-19
- https://doi.org/10.3109/14017438509102815
Abstract
The fatigue endurance limit for Haynes 25 from which the Björk-Shiley (B-S) valve is made is 4900 kg per square centimeter (70000 pounds per square inch). A welded strut may have a variation of fatigue endurance limit from 4900 kg per square centimeter (70000 pounds per square inch) in an incomplete weld with phase segregation. To avoid a weld with this variation, the monostrut has been developed, where the valve is made by electrochemical machining from one piece of Haynes 25. The homogenous outflow monostrut will then have a stable fatigue endurance limit of 4900 kg per square centimeter (70000 pounds per square inch). In a pulse duplicator at a pulse rate of 120 and a peak ventricular pressure of 200 mmHg, the maximum inservice tensile stresses for the 29 mm valve outflow strut do not exceed 700 kg per square centimeter (9800 pounds per square inch). Therefore, a seven times safety limit is present for the monostrut under maximal physiological stress conditions. The struts in the B-S monostrut valve cannot be fractured even under extreme experimental conditions and never under peak physiological conditions.This publication has 9 references indexed in Scilit:
- Optimal Orientation of the 60° and the 70° Björk-Shiley Tilting Disc ValvesScandinavian Journal of Thoracic and Cardiovascular Surgery, 1982
- The Optimal Opening Angle of the Björk-Shiley Tilting Disc Valve ProsthesisScandinavian Journal of Thoracic and Cardiovascular Surgery, 1981
- Ten years’ experience with the Björk-Shiley tilting disc valveThe Journal of Thoracic and Cardiovascular Surgery, 1979
- Radiopaque marker in the tilting disc of the Björk-Shiley heart valveThe Journal of Thoracic and Cardiovascular Surgery, 1977
- RADIOPAQUE MARKER IN TILTING DISK OF BJORK-SHILEY HEART-VALVE - EVALUATION OF INVIVO PROSTHETIC VALVE FUNCTION BY CINE-RADIOGRAPHY1977
- A New Tilting Disc Valve ProsthesisScandinavian Journal of Thoracic and Cardiovascular Surgery, 1969