Acoustic Assessment of the Physical Integrity of Bjo¨rk-Shiley Convexo-Concave Heart Valves

Abstract
Background Several lines of evidence indicate a two-stage failure mode for the Bjo¨rk-Shiley convexo-concave (C/C) heart valve, in which one of the two outlet strut legs separates from the flange before the other, potentially providing an opportunity to identify and prophylactically replace failure-prone valves. Radiographic single leg separation (SLS) detection, although successful, is subjective and skill intensive, implying a need for both an objective preliminary screen and subsequent corroboration of the radiographic findings. Methods and Results We developed a time-windowed, power density analysis of C/C valve closing sounds to detect the vibrational resonance that characterizes the presence of an intact outlet strut in clinically functioning, 29-mm-flange size C/C valves. Recordings from more than 800 patients enrolled in radiographic SLS detection studies were analyzed, and the assessment algorithm was evaluated through a blinded test of 32 study valves for which the true status became known consequent to an autopsy or surgical explantation. Valves were objectively scored on a 0-to-1 scale, with 1 being assuredly intact and scores of Conclusions This level of accuracy is sufficient to serve as an effective preliminary screen, potentially allowing a threefold concentration of SLS prevalence among the C/C valves of patients undergoing radiographic assessment. The value of acoustic classification in avoiding unnecessary operations prompted by false-positive radiographs is less certain.